Interfragmentary Modulus
- 1 Department of Mechanical Engineering, Faculty of Engineering at Si Racha, Kasetsart University, Thailand
Abstract
Problem statement: Young’s modulus and interfragmentary strain (εIF) are the important parameters in bone-plate testing and design. But Young’s modulus is not suitable when we use with εIF in the bone-plate testing. Approach: This research proposed a study on the Interfragmentary Modulus (IM) as the new parameter for the bone-plate testing. Results: The results from the FEM show that the relations of the interfragmentary stress (σIF) and interfragmentary strain are linear for all cases with the condition of the DCP and the screws material properties are linear and isotropic materials. Conclusion: IM are the constant values and we can use IM in the bone-plate testing.
DOI: https://doi.org/10.3844/ajassp.2012.372.375
Copyright: © 2012 Boonthum Wongchai. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
- 3,292 Views
- 3,327 Downloads
- 0 Citations
Download
Keywords
- Interfragmentary modulus
- interfragmentary strain
- interfragmentary stress
- dynamic compression plate
- bone fracture